YOKOGAWA WT1800 User Manual page 82

Precision power analyzer communication interface
Hide thumbs Also See for WT1800:
Table of Contents

Advertisement

5.11 HSPeed Group
:HSPeed:DISPlay:PAGE
Function
Sets or queries the display page of high speed
data capturing mode.
:HSPeed:DISPlay:PAGE {<NRf>}
Syntax
:HSPeed:DISPlay:PAGE?
<NRf> = 1 or 2 (page number)
<NRf> = 1 to 4; on models with the
motor evaluation function (/MTR)
option or auxiliary input (/AUX)
option
:HSPEED:DISPLAY:PAGE 1
Example
:HSPEED:DISPLAY:PAGE? ->
:HSPEED:DISPLAY:PAGE 1
:HSPeed:DISPlay:POVer
Function
Sets or queries the on/off status of the display of
peak over-range information in high speed data
capturing mode.
:HSPeed:DISPlay:POVer {<Boolean>}
Syntax
:HSPeed:DISPlay:POVer?
:HSPEED:DISPLAY:POVER OFF
Example
:HSPEED:DISPLAY:POVER? ->
:HSPEED:DISPLAY:POVER 0
:HSPeed:EXTSync
Function
Sets or queries the on/off status of the high speed
data capturing's external synchronization signal.
:HSPeed:EXTSync {<Boolean>}
Syntax
:HSPeed:EXTSync?
:HSPEED:EXTSYNC OFF
Example
:HSPEED:EXTSYNC ? ->
:HSPEED:EXTSYNC 0
:HSPeed:FILTer?
Function
Queries all high speed data capturing filter
settings.
:HSPeed:FILTer?
Syntax
:HSPeed:FILTer[:HS]
Function
Sets or queries the high speed data capturing
digital filter (HS Filter).
:HSPeed:FILTer[:HS] {OFF|
Syntax
<Frequency>}
:HSPeed:FILTer:HS?
OFF = Digital filter off
<Frequency> = 1 Hz to 1000 Hz (when
the digital filter is on; cutoff
frequency)
:HSPEED:FILTER:HS 100HZ
Example
:HSPEED:FILTER:HS? ->
:HSPEED:FILTER:HS 100.0E+00
:HSPeed:FILTer:LINE?
Function
Queries all high speed data capturing line filter
settings.
:HSPeed:FILTer:LINE?
Syntax
5-50
:HSPeed:FILTer:LINE[:ALL]
Function
Sets the line filters of all the high speed data
capturing elements.
:HSPeed:FILTer:LINE
Syntax
[:ALL] {<Frequency>}
<Frequency> = 0.1 kHz to 100.0 kHz,
300 kHz (cutoff frequency)
:HSPEED:FILTER:LINE:ALL 300KHZ
Example
Description • Line filters are always on in high speed data
capturing mode.
• This command sets the cutoff frequency. You
can set the frequency between 0.1 kHz and
100.0 kHz with a resolution of 0.1 kHz.
:HSPeed:FILTer:LINE:ELEMent<x>
Function
Sets or queries the line filter of the specified high
speed data capturing element.
:HSPeed:FILTer:LINE:
Syntax
ELEMent<x> {<Frequency>}
:HSPeed:FILTer:LINE:ELEMent<x>?
<x> = 1 to 6 (element)
<Frequency> = 0.1 kHz to 100.0 kHz,
300 kHz (cutoff frequency)
:HSPEED:FILTER:LINE:ELEMENT1 300KHZ
Example
:HSPEED:FILTER:LINE:ELEMENT1? ->
:HSPEED:FILTER:LINE:
ELEMENT1 300.000E+03
Description • Line filters are always on in high speed data
capturing mode.
• This command sets the cutoff frequency. You
can set the frequency between 0.1 kHz and
100.0 kHz with a resolution of 0.1 kHz.
:HSPeed:MAXCount?
Function
Sets or queries the maximum number of data
captures.
:HSPeed:MAXCount?
Syntax
:HSPEED:MAXCOUNT? -> 506811
Example
:HSPeed:MEASuring?
Function
Queries all high speed data capturing voltage
mode or current mode settings.
:HSPeed:MEASuring?
Syntax
:HSPeed:MEASuring[:ALL]
Function
Sets all voltage and current modes at the same
time.
:HSPeed:MEASuring[:ALL] {RMS|MEAN|
Syntax
DC|RMEAN}
:HSPEED:MEASURING:ALL RMS
Example
IM WT1801-17EN