To Test The Cables - Agilent Technologies 16962A Service Manual

Logic analyzer
Table of Contents

Advertisement

To test the cables

16962A Logic Analyzer Service Guide
This test allows you to functionally verify logic analyzer
cables and Agilent E5379A probes.
Table 10 Equipment Required to Test Cables
Equipment
Critical Specification
Stimulus Board
No Substitute
Differential Probes
No Substitute
Connect the logic analyzer to the stimulus board.
1
Connect the Agilent E5379A 100- pin differential probes
a
to the logic analyzer cable (also called "Pods") to be
tested. Start with Pods 1 and 2.
Connect the E5379A probe from logic analyzer Pod 1
b
to connector "Pod 4" on the stimulus board.
Connect the E5379A probe from logic analyzer Pod 2
c
to connector "Pod 5" on the stimulus board.
Connect the stimulus board power supply output to the
d
stimulus board power supply connector J82.
e
Plug in the stimulus power supply to line power. The
green LED DS1 should illuminate showing that the
stimulus board is active.
5
Troubleshooting
Recommended Part
16760-60001
E5379A (Qty 2)
71