Set the IP addresses on two analyzers. he addresses can be set to anything, but they must
be di erent. Make sure the subnet mask and gateway addresses are set to
the LAN is active (
LAN tate ON
Connect the two analyzers together using either a crossover cable or a crossover hub.
Now follow the steps in \How to Ping" in order to have the rst analyzer ping the second
analyzer. When done, repeat the procedure having the second analyzer ping the rst. If
both procedures function properly, the LAN circuitry on both analyzers is veri ed. If neither
function, then the following could be wrong:
One or both IP addresses could be wrong.
One or both LAN states could be set to OFF.
he crossover cable could be miswired.
One or both analyzers could be defective.
If possible, eliminate the possibility of a defective analyzer by substitution of a known working
unit. Once the analyzer has been proven to be working properly, concentration can be placed
on the network itself.

Available HP LAN Cables

Cable
Description
HP 92268A
4{meter with attached 8{pin connectors
HP 92268B
8{meter with attached 8{pin connectors
HP 92268C
16{meter with attached 8{pin connectors
HP 92268D
32{meter with attached 8{pin connectors
) on both analyzers. If necessary, refer to \How to Ping."