Keysight Technologies N9030B Service Manual page 98

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
–312
PUD memory lost
–313
Calibration memory
lost
–314
Save/recall memory
lost
–315
Configuration memory
lost
–320
Storage fault;
–321
Out of memory
–330
Self-test failed
–340
Calibration failed
–340
Calibration failed;
<failure msg> <port>
–350
Queue overflow
–360
Communication error
–360
Communication error;
SNS data read
failure. Disconnect
then reconnect SNS
–361
Parity error in
program message
–362
Framing error in
program message
–363
Input buffer overrun
–365
Time out error
98
Verbose/Correction Information
Protected user data saved by the *PUD command has been lost.
The nonvolatile calibration data used by the *CAL? command has been lost.
The nonvolatile data saved by the *SAV? command has been lost.
The nonvolatile configuration data saved by the instrument has been lost.
A problem was found while using data storage. The error is not an
indication of physical damage or failure of any mass storage element.
An internal operation needed more memory than was available. Report this
error to the nearest Keysight Technologies sales or service office.
A self-test failure occurred. Report this error to the nearest Keysight
Technologies sales or service office.
The instrument requires an Align All Now. Restore the alignment by
pressing System, Alignments, Align All Now.
The calibration for one of the I-Q ports did not succeed. The information in
the "failure msg" field can be used to troubleshoot this problem. Contact
Keysight technical support if necessary.
An error occurred that did not get put in the error queue because the
queue was full.
There was a problem with instrument remote communications. The exact
problem cannot be specifically identified.
The Keysight Smart Noise Source connected to the PXA has failed to be
read by the application. Please disconnect and reconnect the SNS. If this
continues to fail, then the SNS may have had its EEPROM corrupted or
another hardware fault exists. Check SNS on another instrument, NFA and
ESA are also SNS compatible instruments. Check the device is not an
Keysight power sensor which uses the same cable interface.
A parity bit was not correct when the data was received. For example, on a
parallel port.
A stop bit was not detected when data was received. For example, on a
remote bus port.
A software or hardware input buffer on a port overflowed with data
because of improper or nonexistent pacing.
There was a time-out problem in the instrument. The exact problem cannot
be specifically identified.