Keysight Technologies N9030B Service Manual page 115

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
–230
Data corrupt or stale;
Unable to load state from
file
–231
Data questionable
–232
Invalid format
–232
Invalid format; Map
information not loaded
–233
Invalid version
–240
Hardware error
–240
Hardware error; See
details in Windows Event
Log under SA
–241
Hardware missing
–241
Hardware missing; Internal
preamp not available at
all frequency points
–241
Hardware missing; not
available for this model
number
–241
Hardware missing; Option
not installed
–250
Mass storage error;
–250
Mass storage error; Access
denied
–250
Mass storage error; Bad
path name
–250
Mass storage error; Can
only import single trace
.csv files
N9030B PXA Signal Analyzer Service Guide
Verbose/Correction Information
There is something wrong with the state data in the desired file.
Maybe the file is corrupt, or it is from an instrument/version that is
not recognized by the current instrument.
Indicates that the measurement accuracy is suspect
A data element was found but it could not be used because the
data format or the data structure was not correct.
Instrument failed to load the burst mapping information from the
selected file.
A legal data element was found but could not be used because
the version of the data is incorrect. For example, state data
changes as new instrument features are added, so old state files
may not work in an instrument with a newer version of software.
A legal program command or query could not be executed
because of a hardware error. The exact problem cannot be
specifically identified.
The internal data acquisition system detected a problem at
startup and logged the details in the Windows Event Log.
The operation could not be performed because of missing
hardware; perhaps the optional hardware is not installed.
The Internal Preamp is currently turned on, but the measurement
is being performed completely or partially outside the range of the
preamp. It is recommended that the user turns preamp off to
ensure consistent results across the entire measurement.
The hardware required is not part of this model
The optional hardware is not installed.
A problem was found with the mass storage device (memory, disk
drive, etc.). The exact problem cannot be specifically identified.
Access is denied.
The specified path is invalid.
Trace files containing multiple traces can not be imported.
However, if you need to recall multiple traces you can use the
Save and Recall functions rather than the Import and Export
functions.
115