Keysight Technologies N9030B Installation Note

Keysight Technologies N9030B Installation Note

Pxa spectrum analyzer, option c35, 3.5 mm input connector retrofit kit
Hide thumbs Also See for N9030B:

Advertisement

Quick Links

Keysight Technologies
N9030B PXA Spectrum Analyzer
Option C35, 3.5 mm Input Connector
Retrofit Kit
Installation Note

Advertisement

Table of Contents
loading

Summary of Contents for Keysight Technologies N9030B

  • Page 1 Keysight Technologies N9030B PXA Spectrum Analyzer Option C35, 3.5 mm Input Connector Retrofit Kit Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 The Option C35 retrofit kit provides all the parts and instructions necessary to install the 3.5 mm input connector into the Keysight N9030B Option 526 (26.5 GHz spectrum analyzer). The latest revision of the X-Series Signal Analyzers software may be downloaded from http://www.keysight.com/find/xseries_software...
  • Page 4 — T-20 Torx driver — T-10 Torx driver — 5/16-inch open end wrench — N9030B PXA Service Guide can be found online at the following URL: www.keysight.com/find/N9030B_service_guide — Personal computer with internet access and USB port — USB storage device with > 2 GB free memory If adjustments and performance testing is required: —...
  • Page 5: Installation Procedure

    Installation Procedure Installation Procedure Remove the Instrument Outer Case, Front Panel Assembly, and Right Side Chassis (RF Bracket) If the instrument is placed on its face during any of the following procedures, be sure to use a soft surface or soft cloth to avoid damage to the front panel, keys, or input connector. See the Instrument Outer Case, Front Frame Assembly and RF Area (RF Bracket) removal procedures in the Service Guide's “Assembly Replacement Procedures”...
  • Page 6: Installation Procedure Over Usb

    Installation Procedure over USB Installation Procedure over USB 1. Locate the Option Upgrade Entitlement Certificate from the kit. 2. Redeem the Option Upgrade Entitlement Certificate by following the instructions on the Certificate. 3. After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an attached License File.
  • Page 7: Verify The Installation

    The License File can be manually installed over USB or LAN by placing the license file in the following folder on the signal analyzer C:\Program Files\Agilent\licensing Verify the Installation 1. Press System, Show System to display a list of installed options. 2. Verify that N9030B-C35 appears on the list. Installation Note N9030-90078...
  • Page 8: Adjustments Required

    For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...