Analog/Digital IF Troubleshooting
A3 Digital IF Troubleshooting
12.The analyzer should read 22.5 MHz at −25 dBm ± 4 dB as shown in
Figure 7-8
22.5 MHz IF Input
13.If the 22.5 MHz signal is not measuring the correct power level, see
14.If the 22.5 MHz signal is within tolerance, carefully reconnect the W42
242
7-8.
Analog IF Troubleshooting" on page 230
cable to A3J19.
in this service guide.