Keysight Technologies N9030B Service Manual page 114

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
–224
Illegal parameter value;
Invalid list length
–224
Illegal parameter value;
LXI Event <event> already
exists.
–224
Illegal parameter value;
LXI Event <event> contains
illegal characters.
–224
Illegal parameter value;
LXI Event <event> does not
exist.
–224
Illegal parameter value;
Measurement not available
–224
Illegal parameter
value;This instrument is
always DC coupled
–224
Illegal parameter
value;This model is always
AC coupled
–225
Out of memory
–225
Out of memory;
Insufficient resources
to load Mode <mode name>
–225
Out of memory; Memory
limit caused Data
Acquisition to be
truncated
–226
List not same length
–230
Data corrupt or stale;
–230
Data corrupt or stale;
Measurement data is not
available
–230
Data corrupt or stale;
Trace contains no data.
114
Verbose/Correction Information
You are trying to set some list measurement settings, but the
multiple lists that you sent were not all the same length. The
number of settings must be consistent from list to list.
This error occurs when you try to add an LXI Event that has
already been added.
When a new LXI Event is created, it may not use the comma,
semicolon, or newline characters. All other printable ASCII
characters are valid.
The requested event has not been added yet.
You tried to turn on a measurement that is not available in the
current mode.
You can't set AC coupling in this instrument
You can't set DC coupling in this analyzer
There is not enough memory to perform the requested operation.
If you attempt to load a mode via SCPI that will exceed memory
capacity, the Mode does not load and this message is returned.
"mode name" is the SCPI parameter for the Mode in question, for
example, SA for Spectrum Analyzer Mode. You can free up
resources in the System, Power On, Configure Applications menu
You are using the LIST structure, but have individual lists that are
not the same lengths.
A legal data element was found, but it could not be used because
the data format or the data structure was not correct. Maybe a
new measurement had been started but had not completed.
Measurement data not available. The measurement that you are
trying to get data from must be the current active measurement.
Maybe you have not initiated the measurement, or it has not
completed all the sweeps/averages needed.
Trace cannot be displayed because currently there is no data
assigned to it. Use the functions under the Trace menu, or load a
previously saved trace, to assign data to the trace.
N9030B PXA Signal Analyzer Service Guide