Instrument Messages
Event Messages
Err#
Message
–221
Settings conflict; Log Scale
Type is not available with
Demod View
–221
Settings conflict; Log Scale
Type is only available in
swept measurement
–221
Settings conflict; Marker 1
Trace Update=off turns off
Signal Track
–221
Settings conflict; Marker
cannot be relative to itself
–221
Settings conflict; Marker
Count is not available when
Gate on
–221
Settings conflict; Marker
Function is not available
for a Fixed marker
–221
Settings conflict; Marker
type must be delta
–221
Settings conflict; Marker->
function is not available in
zero span
–221
Settings conflict; Mask
unavailable for current
Span. Increase to display
mask.
–221
Settings conflict; Meas Type
was changed to Examine for
Exp Avg Mode.
–221
Settings conflict; Meas Type
was changed to Full for
Repeat Avg Mode
–221
Settings Conflict;
MinPts/RBW limit not met
–221
Settings conflict; Mkr -> CF
is not available when the
x-axis is time domain
–221
Settings conflict; MS gain
is not available in this
Mode
104
Verbose/Correction Information
The logarithmic x-axis scales are not available when you
have the demod view turned on.
Logarithmic scaling can be used when making a swept SA
measurement. It is not available in the SA measurement
when you are using FFT sweeps.
Signal Track not available unless the trace containing Marker
1 is updating
A marker must be set relative to another marker, not to
itself.
The marker count function cannot be used while you have
gating turned on.
If a Marker is a Fixed type marker, the marker's value does
not change from when it first became fixed. You cannot turn
on or change a Marker Function because there is no ongoing
measurement data to use for the marker function
calculation.
MkrΔ->Span and MkrΔ->CF require that the selected
marker be a delta marker.
Most of the "Marker To" functions are not available if you are
in zero span (span = 0 Hz, or time domain). So you cannot
send the commands for these functions.
The current span setting is either narrower than the mask
width or so wide that there are too few display points to
allow the mask to be drawn. Increase or decrease the span
to display the mask.
Average Mode has been changed to Exponential. Full Meas
Type is not available for Exponential Average Mode therefore
Meas Type has been changed to Examine.
Average Mode has been changed to Repeat. Examine Meas
Type is not available for Repeat Average Mode therefore
Meas Type has been changed to Full.
The marker to center frequency functionality does not work
when the x-axis is in the time domain.
Mobile Station gain correction is not available in some
Modes, or in some measurements (for example, the SA
measurement).