Keysight Technologies N9030B Service Manual page 577

Pxa signal analyzer
Table of Contents

Advertisement

Functional Tests
Test Equipment
Table 18-1
Analyzer
Option
Item
Preamp
Low Power Sensor
other than
P03
Miscellaneous Equipment
All
Filter, 50 MHz Low Pass
544, 550
Power Splitter, 2.4 mm
526
Power Splitter, 3.5 mm
503, 508,
Power Splitter
513
All
Termination, 50 W
BBA
Termination, 50 W
a. Maximum frequency equal to or greater than the maximum frequency of the PXA.
b. PSG model numbers: E8244A, E8254A, E8247C Option 520, E8247C Option H31 or 540, E8257D
Option 520, E8257D Option 550, E8257D Option 567, E8267D Option 544.
c. Maximum frequency equal to or greater than the maximum frequency of the preamplifier option.
N9030B PXA Signal Analyzer Service Guide
Critical Specifications
Frequency Range: 50 MHz to 50 GHz
Amplitude Range: –70 to –20 dBm
Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Rejection at 80 MHz: > 60 dB
Frequency: 10 MHz to 26.5 GHz
Nominal Insertion Loss: 8.5 dB
Tracking Between Ports: < 0.4 dB
Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
Type N (m) Connector
Frequency: 30 Hz to 26.5 GHz
BNC (m) Connector
Frequency: 250 kHz to 40 MHz
Al ternate
Recommended
Keysight
Keysight Model
Model
8487D
8485D c
c
0955-0306
11667C
11667B
11667A
11667B
909A
Option 012
11593A
577