Instrument Messages
Event Messages
Err#
Message
–250
Mass storage error; Can
only load an Antenna Unit
into Correction 1
–250
Mass storage error; Cannot
make
–250
Mass storage error;
Different Antenna Unit
already in use
–250
Mass storage error;
Directory not found
–250
Mass storage error; Failed
to Load trace. Bad file
format.
–250
Mass storage error; File
<filename> wrong type
–250
Mass storage error; File
<filename> and instrument
version mismatch
–250
Mass storage error; File
contains incorrect data
for this operation
–250
Mass storage error; File
empty
–250
Mass storage error;
Invalid register number
for *SAV or *RCL Mass
Storage error
–250
Mass storage error; Lock
violation
–250
Mass storage error; Mkr
Table must be on to save
Mkr Table as Meas Results
–250
Mass storage error; No
file names available
–250
Mass storage error; Open
failed
116
Verbose/Correction Information
The only Correction register that supports Antenna Units is
number 1. You have attempted to load an Ampcor file which
contains antenna units into another register
The directory or file cannot be created.
Attempt to import Corrections file with Antenna Unit that differs
from an in-use correction.
The system cannot find the path specified.
The load trace operation could not be completed, as the input file
was not in the expected format. You can only load traces that were
previously saved using the 'Save Trace' feature.
Attempt to import a data file that is not the proper type for this
operation.
While opening a file, there was a mismatch between file version or
model number with instrument version or model number. The
import still tried to load as much as possible, but you should
check it closely.
There is a mismatch between the file data type of the file specified
and the destination indicated. For example, a correction set
cannot be loaded/imported into a limit line.
Cannot save trace because it contains no data. Check that the
trace is turned on and contains some valid data.
You have used the *SAV command to save a state to a non
existent state register.
Or
You have used the *RCL command to recall a state register that
wasn't previously saved with the *SAV command.
The process cannot access the file because another process has
locked a portion of the file.
You have to have a Marker Table on the screen before you can
save it. Turn on the Marker Table and try again.
Attempt to use the auto file name generation when all 10,000 file
names are taken.
The system cannot open the device or file specified. This could be
because the storage media is full, or possibly due to a filename
error. If using an external storage devise, check that the devise is
properly formatted.
N9030B PXA Signal Analyzer Service Guide