Instrument Messages
Event Messages
Err#
Message
–221
Settings conflict;
Destination trace for Trace
Math cannot be a trace
operand
–221
Settings conflict; Downconv
only available when DUT is
Amplifier
–221
Settings conflict; EDGE EVM
only supports EDGE TCH burst
type.
–221
Settings conflict;
Electronic attenuator is
disabled
–221
Settings conflict;
Electronic attenuator is not
available above 3.6 GHz
–221
Settings conflict;
Electronic attenuator
unavailable in current state
–221
Settings conflict;
Electronic attenuator
unavailable with Preamp on
–221
Settings Conflict; FAST
method can only be used
while Radio Std is W-CDMA
–221
Settings conflict; Feature
not available in this View
–221
Settings conflict; Feature
not supported for selected
source
–221
Settings conflict; Feature
not supported for this
measurement.
–221
Settings conflict; Feature
not supported for this model
number
N9030B PXA Signal Analyzer Service Guide
Verbose/Correction Information
The resulting trace data (from doing a trace math function)
cannot be put into the any of the traces that are being used
by the math operation.
SCPI only message. The System Downconverter can only be
set to 'On' when the DUT type is amplifier. Change DUT type
to Amplifier if the System Downconverter is required.
You are using the mechanical attenuator, and have not
enabled the electronic attenuator. You cannot set the value
of the electronic attenuator because it automatically
sets/changes when enabled.
The maximum frequency of the electronic attenuator is 3.6
GHz. This is because of switching capacitance.
The internal preamp is on. Electronic attenuator cannot be
used while you are using the internal preamp.
Some functionality is available in one View, but not in
another. (See the Views under the View/Display key.) This
error occurs if you send a SCPI command or push a gray-out
key that is not available in the current selected View.
You have asked for a feature that the selected source is not
capable of.
Some functionality is available in one measurement, but not
in another. (See the measurements under the Meas key.)
This error occurs if you send a SCPI command or push a
gray-out key that is not available in the current selected
measurement.
This functionality is not a part of the instrument you are
using but may be found in other models in the X-Series.
101