RF Area (Options 503, 508, 513, 526)
8. Refer to
Figure 16-33
Bracket Mount to Switch
N9030B PXA Signal Analyzer Service Guide
Figure
16-33. Install the bracket (1) to the switch (2) using three
screws (3) (0515-1934). Assure the number 3 on the switch is positioned
as shown. Torque screws to 6 inch-lbs. Install the ribbon cable (4) into the
switch connector header.