Keysight Technologies N9030B Service Manual page 592

Pxa signal analyzer
Table of Contents

Advertisement

Functional Tests
Frequency Response (Flatness)
Table 18-7
Frequency Response (Flatness) Results
Analyzer
Amplitude
Center
Frequency
Meas
amp
50 MHz
1 GHz
2 GHz
3 GHz
3.5 GHz
4 GHz
6 GHz
8 GHz
9 GHz
11 GHz
13 GHz
14 GHz
17 GHz
20 GHz
23 GHz
26 GHz
29 GHz
32 GHz
35 GHz
38 GHz
41 GHz
42.5 GHz
43.5 GHz
46.5 GHz
49.5 GHz
592
Power Meter
Meas Error
Measurement
Power
Meas
meter
error
Flatness
Normalized
to 50 MHz
Flat
Norm
0
N9030B PXA Signal Analyzer Service Guide
Flatness Error
Test Limits
(dB)
Ref
±1.5 dB
1.5 dB
±1.5 dB
±1.5 dB
±2.5 dB
±2.5 dB
±2.5 dB
±3.0 dB
±3.0 dB
±3.0 dB
±3.5 dB
±3.5 dB
±3.5 dB
±3.5 dB
±3.5 dB
±3.5 dB
±3.5 dB
±4.2 dB
±4.2 dB
±4.2 dB
±4.2 dB
±4.2 dB
±4.2 dB
±4.2 dB

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents