Keysight Technologies N9030B Service Manual page 112

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
–201
Invalid while in local
–202
Settings lost due to rtl
–203
Command protected
–203
Command protected; feature
not licensed
–210
Trigger error
–211
Trigger ignored
–212
Arm ignored
–213
Init ignored
–214
Trigger deadlock
–215
Arm deadlock
–220
Parameter error
–221
Settings conflict;
–222
Data out of range;
–222
Data out of Range; clipped
to source max/min
–222
Data out of range; Invalid
list data
112
Verbose/Correction Information
The command cannot be executed while the instrument in Local
control.
A "return to local" control was forced and some settings were lost
as a result of this.
The command could not be executed because it is disabled. It
was disabled by licensing or password protection
The specified feature, for example "N9073A-TR2" is not licensed.
The license may have expired. You cannot use it until you get a
license.
A trigger error has occurred, but the exact problem cannot be
specifically identified.
A GET, *TRG or other triggering signal was received, but it was
ignored because of timing considerations. For example, maybe
the instrument was not ready to respond when the command was
received.
An arming signal was received, but it was ignored.
An initiate trigger/sweep request was received and ignored,
because another measurement was already in progress.
The trigger source for the initiation of a measurement is set to
GET, and the following measurement query was received. The
measurement cannot be started until a GET is received, but the
GET would cause an INTERRUPTED error.
The arm source for the initiation of a measurement is set to GET
and the following measurement query is received. The
measurement cannot be started until a GET is received and the
GET would cause an INTERRUPTED error.
A problem was found with a program data element. The exact
problem cannot be specifically identified.
There are many types of settings conflict errors. See section 3.5
for information about these errors.
A data element was found but the instrument could not be set to
that value because it was outside the range defined for the
command. A descriptive message may be appended, such as
"clipped to upper limit"
A source parameter has been entered that exceeds the range of
the selected source. The parameter has been clipped to match
the range of the source
You tried to use a trace that has a number of sweep points that is
different from the current setting of sweep points.
N9030B PXA Signal Analyzer Service Guide