Keysight Technologies N9030B Installation Note

Pxa signal analyzer, option hl1 microwave preselector bypass (mpb) and low noise path (lnp)
Hide thumbs Also See for N9030B:

Advertisement

Quick Links

Keysight Technologies
N9030B PXA Signal Analyzer
Option HL1 Microwave Preselector Bypass
(MPB) and Low Noise Path (LNP)
Installation Note

Advertisement

Table of Contents
loading

Summary of Contents for Keysight Technologies N9030B

  • Page 1 Keysight Technologies N9030B PXA Signal Analyzer Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit Products Affected: PXA N9030B, Option 508, 513, and 526 Serial Numbers: To Be Performed By: (X) Keysight Service Center...
  • Page 4: Installation Kit Parts List

    Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit Installation Kit Parts List Quantity Description Keysight Part Number Installation Note This note Option Upgrade Entitlement Certificate 5964-5178 Coax Switch N1810-60069 Transfer Switch, (top switch) 87222-60026 Coaxial Fixed Attenuator 08493-60026 Bracket, Switch E4410-00104...
  • Page 5: Tools Required

    Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit Tools Required — T-8 TORX Driver — T-10 TORX Driver — T-20 TORX Driver — 5/16-inch torque wrench — Keysight Calibration and Adjustment Software, N7814A (revision E.16.00 or later) —...
  • Page 6: Installation Procedure

    Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit Installation Procedure If the instrument is placed on its face during any of the following procedures, be sure to use a soft surface or soft cloth to avoid damage to the front panel, keys, or input connector. Make sure any adapters on the front panel are removed.
  • Page 7 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 2. Locate the YTF and remove W56 from between the YTF and the Low Band Switch Assembly, and remove W57 between the YTF and the Front End Assembly at J9. These cables will not be reused.
  • Page 8 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 3. Refer to Figure 3. Locate the other N1810-60069 switch in the kit, and place switch (1) on the bracket previously installed. Assure Keysight label on switch is facing up. Figure 3 Switch 2 Placement 4.
  • Page 9 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 7. Refer to Figure 4. Route W7 over bracket and switch as shown and insert W7 connector into Low Band Switch header Figure 4 W7 Routing Installation Note N9030-90083...
  • Page 10 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 8. Refer to Figure 5. Locate Wire Harness E4410-60159 in the kit and attach the end with the single connector to the Front End Control assembly J801 connector. Route the cable between the inner chassis and the switches just installed, and mate with connectors (1) on the switches.
  • Page 11 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 9. Refer to Figure 6. Attach the coaxial fixed attenuator (1) from the kit to switch port 1. Torque to 10 inch-lbs. Figure 6 W7 Routing Installation Note N9030-90083...
  • Page 12 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 10.Refer to Figure 7. Locate the 87222-60026 switch, N9020-00022 bracket, and N9020-60087 ribbon cable in the kit. Install the bracket (1) to the switch (2) using three 0515-1934 screws provided (3).
  • Page 13 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 11.Install the switch assembly (1) as shown in Figure 8. Route the switch control cable between the switch and the chassis. Connect the ribbon cable to the Front End Controller J802 as shown.
  • Page 14 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 12.Refer to Figure 9. Install rigid cable N9020-20146 (1) between top switch port 4 and Low Band Switch Assembly connector that is closest to the inner chassis. Torque to 10 inch-lbs. Figure 9 N9020-20146 Rigid Cable Installation Installation Note N9030-90083...
  • Page 15 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 13.Refer to Figure 10. Install rigid cable N9020-20210 (1) between attenuator (connected to center switch port 1 and bottom switch port 1. Torque to 10 inch-lbs. Figure 10 N9020-20210, N9020-20144, and N9020-20164 Rigid Cables Installation 14.Install rigid cable N9020-20144 (2) from top switch port 2 to center switch center port.
  • Page 16 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 16.Refer to Figure 11. Install rigid cable N9020-20145 (1) from top switch port 3 to Low Band Switch Assembly as shown. Torque to 10 inch-lbs. Figure 11 N9020-20145, N9020-20165, and N9020-20163 Rigid Cables Installation 17.Install rigid cable N9020-20165 (2) from bottom switch port 2 to YTF Assembly as shown.
  • Page 17 Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit 19.Refer to Figure 12. Install rigid cable N9020-20143 (1) from top switch port 1 to Attenuator B output as shown. Torque to 10 inch-lbs. Figure 12 N9020-20143 Rigid Cable Installation 20.Replace the right side chassis, front panel, top brace, and instrument cover.
  • Page 18: License Installation Procedure Over Usb

    Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit License Installation Procedure over USB Even though the retrofit kit was ordered as Option HL1, the kit installs instrument options MPB and LNP. The license file mentioned below will install both option MPB and LNP. 1.
  • Page 19: Verify The Installation

    Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit Figure 14 Successful License Installation Verify the Installation 1. Cycle the power on the signal analyzer. 2. Press System, Show System to display a list of installed options. 3.
  • Page 20 For assistance, get in touch with your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...
  • Page 21 This information is subject to change without notice. © Keysight Technologies 2016 Edition 1, February 2016 N9030-90083 www.keysight.com...