Keysight Technologies N9030B Service Manual page 36

Pxa signal analyzer
Table of Contents

Advertisement

Overview
Service Equipment You Will Need
Table 1-1
Instrument
Low Power Microwave
Power Sensor
(2 required)
(for Option P08, P13,
or P26 only)
Power Sensor Cable
(2 required)
Standard s
Frequency Standard
50 MHz,
−25 dBm Calibrator
Attenuators
10 dB
Step Attenuator
1 dB
Step Attenuator
Attenuator
Interconnect Kit
Attenuator Driver
36
Required Test Equipment
Critical Specifications
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: −70 to +20 dBm
Zero Set: ± 50 pW
Zero Drift: < ± 4 pW
Measurement Noise: < 4 pW
Cal Factor Uncertainty: < 3.0%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
Input Connector: 3.5 mm (m)
Frequency: 10 MHz
f
Accuracy: < ±1 e10
Frequency Drift: < 2.5 kHz
Typical VSWR: 1.06:1
Output Power Variation: ±.004 dB
Total Harmonic Content: −45 dBc
Range: 0 to 80 dB
Accuracy:
Characterized by standards lab
VSWR: at 50 MHz: ≤ 1.05:1
Range: 0 to 9 dB
Accuracy:
Characterized by standards lab
VSWR: at 50 MHz: ≤ 1.05:1
Type N connector kit to connect 8496G
to 8494G
Compatible with the 8496G and 8494G
step attenuators.
Recommended
Keysight Model
a
Number
or Vendor
Model Number
g
8485D
11730A
Symmetricom 5071A Keysight
Z5602A or B
Opt H51 for Type-N
models
Opt H35 for 3.5 mm
models
Opt H24 for 2.4
models
8496G
(Option 001, H50)
h
8494G
(Option 001, H50)
h
11716A
11713B
N9030B PXA Signal Analyzer Service Guide
Al ternative
b
Use
Keysight Model
a
Number
A, P
A, P, T
A, P
5061B, 5071A
A
8496H
P
(Option 001,
H50)
8494H
P
(Option 001,
H50)
P
11713A
P

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents