Keysight Technologies N9030B Service Manual page 133

Pxa signal analyzer
Table of Contents

Advertisement

Instrument Messages
Condition Messages
Err#
Bit in
Status
Register
207
2
207
2
N9030B PXA Signal Analyzer Service Guide
Message
Burst Not Found
Burst not found;with
selected Time Slot
Error or
More Information
Warning
E
The burst signal cannot be detected because
of inappropriate parameter settings or
incorrect signal.
An in appropriate parameter setting could
cause the signal to be partially, rather than
fully, on the display, Burst Search Threshold
and/or Burst Search Length may need to be
adjusted.
An incorrect signal could have either
insufficient power, the rising or falling edges
cannot be detected, or the burst is less than
126 microseconds.
Carrier signal is not actually bursted.
W-CDMA: Either the signal being analyzed has
insufficient power, the rising or falling edges
cannot be detected, or the burst is less than
126 microseconds.
W_CDMA: Cannot synchronize measurement
with PRACH channel for Power Control
measurement, because the signal cannot be
found. Make sure PRACH is present in the
W-CDMA uplink signal, and that the preamble
signature and scramble code are set correctly.
GSM: Data was acquired but a GSM burst was
not found, with the timeslot mode disabled.
NADC, PDC: A valid burst is not found when
the Device is MS.
1xEV-DO: Data was acquired but a 1xEV burst
was not found, with the timeslot mode
disabled.
Bluetooth: The burst that has been found
does not correspond to the currently selected
Bluetooth packet type (the burst length may
be too short).
WLAN: The instrument cannot find a valid
WLAN burst. You may need to extend the
search length.
In the PSA, this error was reported as one of
the following error numbers: 10772, 13104,
10160, 10286, 10420, 10454, 10614, 10904,
10928, 13074, 10287
E
The selected timeslot does not contain the
expected burst.
133

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents