Assembly Replacement Procedures
RF Area (Options 503, 508, 513, 526)
18.Refer to
Figure 16-41
Ribbon Cable Routing
19.Refer to
20.Replace the front panel. Refer to the
21.Replace the instrument outer case. Refer to the
472
Figure
16-41. Route all ribbon cables as shown to avoid being
damaged when the top brace is installed.
Figure
16-7. Position the RF bracket onto the chassis and replace
the sixteen screws (0515-0372). Torque to 9 inch-pounds.
procedure.
replacement procedure.
Front Frame Assembly
Instrument Outer Case
N9030B PXA Signal Analyzer Service Guide
replacement