Instrument Messages
Condition Messages
Condition Errors 67 to 95, Calibration Failure (Extended)
This series of errors corresponds to the bits in the
STATus:QUEStionable:CALibration:EXTended:FAILure sub-register. The second
column in the table shows the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 71 indicates that the Characterize Preselector Failure has
been detected, error 1071 indicates that failure has been cleared.
This register is summarized as bit 9 of the STATus:QUEStionable:CALibration
register. See section
on page
Err#
Bit in
status
register
67
0
69
1
71
2
73
3
75
4
77
5
79
6
81
7
83
8
85
9
87
10
89
11
91
12
93
13
95
14
N9030B PXA Signal Analyzer Service Guide
"Condition Errors 36 to 64, Calibration Needed or Failed"
124.
Message
Align 9kHz to 30MHz
failed
Align 30MHz to 1GHz
failed
Characterize Preselector
failure
unused
unused
unused
unused
unused
unused
unused
unused
unused
unused
unused
unused
Error or
More Information
Warning
W
On PSA, this was error 13749
W
On PSA, this was error 13751
W
The preselector characterization routine
failed.
127

Advertisement

Table of Contents

Troubleshooting