Instrument Messages
Condition Messages
Condition Errors 65 to 92, Calibration Needed (Extended)
This series of errors corresponds to the bits in the
STATus:QUEStionable:CALibration:EXTended:NEEDed sub-register. The
second column in the table shows the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 72 indicates that the "Input Attenuation not calibrated"
condition has been detected, and error 1072 indicates that failure has been
cleared.
This register is summarized as bit 8 of the STATus:QUEStionable:CALibration
register. See section
on page
Err#
Bit in
status
register
3
65
0
66
1
68
2
72
4
74
5
76
6
78
7
80
8
82
9
84
10
86
11
88
12
90
13
92
14
126
"Condition Errors 36 to 64, Calibration Needed or Failed"
124.
Message
System alignment required
unused
Align 9kHz–30MHz required
Align 30MHz–1GHz required
Input Attenuation not
calibrated
unused
unused
unused
unused
unused
unused
unused
unused
unused
unused
Error or
More Information
Warning
E
An EMI conducted frequency range
alignment is needed.
E
An EMI radiated frequency range
alignment is needed.
E
Corrected measurements have been
requested and the required RF
front-end setting of x dB has not been
calibrated.
N9030B PXA Signal Analyzer Service Guide