Post-Repair Procedures
Post-Repair Procedures
Data Restore
1. Connect a USB mouse to one of the instrument front panel USB ports.
2. Locate the USB FLASH drive that was earlier used to backup the
3. Restore the instrument calibration data by pressing: System, Alignments,
4. Follow the on-screen instruction to restore the previously saved
5. Cycle instrument power.
560
instrument calibration file and insert it into one of the instrument front
panel USB ports.
Backup or Restore Align Data... The Alignment Data Wizard window as
shown in
Figure 17-1
calibration data from the USB FLASH drive.
will appear.
N9030B PXA Signal Analyzer Service Guide