Instrument Messages
Advisory Messages
Message
Span is not coupled to RBW
when EMI detector is selected
Sweep Points/Span is <
minimum. Results may be
inaccurate.
Sweep Setup is not available
in Zero Span
Sync is RF Ampl (not Training
Sequence). Bits are not
accurate.
Trace file saved.
Use Gate View Sweep Time in
the Gate menu.
User Cal valid. Apply Cal
from Meas Setup menu
N9030B PXA Signal Analyzer Service Guide
Description/Correction Information
The sweep point to span ratio is below the minimum required to ensure the
bucket ratio is large enough to test DVB-T masks
Zero span is a display at a single frequency, so there is no "sweeping" to
set up.
The trace saving operation was successful.
When in Gate View you use Gate View Sweep Time, rather than Sweep
Time, to control the Gate View window
The measurement setup has changed such that the current cal data can be
applied to the results. To apply the cal, press Meas Setup/Cal Setup/Apply
Calibration. A new cal can be performed if required.