Assembly Replacement Procedures
Front Frame Assembly
Front Panel Shield
Figure 16-102
Front Panel Shield
1. Refer to
2. To remove the front panel shield (1), remove the eight screws (2)
3. When replacing the front panel shield, torque all screws to 9 inch-pounds,
544
Figure
16-102. Remove ribbon cable W1.
(0515-0372).
When replacing ribbon cable W1 press firmly on each side to ensure the
hooks catch on the boards mating connector.
N9030B PXA Signal Analyzer Service Guide