Instrument Messages
Condition Messages
Err#
Bit in
Message
status
register
159
13
Settings Alert; Analog Out
settings conflict
159
13
Settings Alert;I/Q
mismatch:<Differential|Input
Z|Attenuation>
159
13
Settings Alert;Parm/data
mismatch
161
14
Setting Modified
161
14
Setting Modified; Filter not
applied
N9030B PXA Signal Analyzer Service Guide
Error or
More Information
Warning
W
The user has manually set the Analog
Output under the Input/Output menu to
a setting that conflicts with the current
measurement. There will be no output
on the Analog Out port until this conflict
is resolved. In most cases, simply set
Analog Out to Auto for the optimal
setting.
W
The impedance, differential, or
attenuation settings for the I and Q
channels do not match. For valid I+jQ
measurements the impedance and
differential settings should be the same
on both channels and the attenuation
should match within 1 dB
W
For Bluetooth, the detected
parameters did not match the data
E
E
The filter you have selected is larger
than the sampling frequency. You
should select a different filter.