Overview
Service Equipment You Will Need
The interconnect cable should NEVER be disconnected once the loss
characterization is performed.
Table 1-2
1 dB Step Attenuator
Nominal Attenuation
Attenuator Section
(dB)
#1
(1 dB)
0
Off
1
On
2
Off
3
On
4
Off
5
On
6
Off
7
On
8
Off
9
On
a. TME requires that Section 3 be characterized.
Table 1-3
10 dB Step Attenuator
Nominal Attenuation
Attenuator Section
(dB)
#1
(10 dB)
0
Off
10
On
20
Off
30
On
40
Off
50
On
60
Off
70
On
a. TME requires that Section 3 be characterized.
N9030B PXA Signal Analyzer Service Guide
#2
#3
(2 dB)
(4 dB)
Off
Off
Off
Off
On
Off
On
Off
Off
a
On
Off
On
On
On
On
On
Off
On
Off
On
#2
#3
(20 dB)
(40 dB)
Off
Off
Off
Off
On
Off
On
Off
Off
a
On
Off
On
On
On
On
On
Recommended Uncertainty
(dB)
#4
(4 dB)
Off
0 (Reference)
Off
< 0.005
Off
< 0.005
Off
< 0.005
Off
< 0.005
Off
< 0.005
Off
< 0.005
Off
< 0.005
On
< 0.005
On
< 0.005
Recommended Uncertainty
(dB)
#4
(40 dB)
Off
0 (Reference)
Off
< 0.010
Off
< 0.015
Off
< 0.020
Off
< 0.025
Off
< 0.030
Off
< 0.035
Off
< 0.040
43