Assembly Replacement Procedures
RF Area (Options 503, 508, 513, 526)
10.Refer to
Figure 16-22
Switch Assembly Removal
11.Refer to
Figure 16-23
Bracket Mount to Switch
N9030B PXA Signal Analyzer Service Guide
Figure
16-22. Disconnect the ribbon cable to the Front End
Controller J802. Remove the switch assembly (1) by removing the two
screws (0515-2032) at the frame (2) and the single screw on the side (3)
(0515-0372).
Figure
16-23. To separate the switch (2) from the bracket (1)
(N9020-00022), remove the three screws (3) (0515-1934).
455