Keysight Technologies N9030B Service Manual page 34

Pxa signal analyzer
Table of Contents

Advertisement

Overview
Service Equipment You Will Need
Table 1-1
Instrument
Function Generator
Counters
Universal Counter
Meters
Digital Multimeter
Power Meter
RF Power Sensor
(2 required)
34
Required Test Equipment
Critical Specifications
Frequency: 3 Hz to 300 kHz
Amplitude Resolution: 0.1 mv
Harmonic Distortion: -35 dBc
Frequency: 10 MHz
Gate time: 10 to 100 seconds
Must be capable of measuring signal at
+7 dBm (0.5 Vrms)
Frequency: 3 Hz to 300 kHz
AC Accuracy: ± 0.31% of reading
Dual Channel
Instrumentation Accuracy: ± 0.5%
Power Reference Accuracy: ± 0.6%
Compatible with 8480 series power
sensors
dB relative mode
Frequency Range: 100 kHz to 3.6 GHz
Amplitude Range: −30 to +20 dBm
Zero Set: ± 50 mW
Zero Drift: < ± 10 nW
Measurement Noise: < 110 nW
VSWR
100 kHz to 1 MHz: ≤ 1.20:1
1 MHz to 2 GHz: ≤ 1.10:1
2 GHz to 3 GHz: ≤ 1.30:1
Cal Factor Uncertainty (std): < 1.6%
Option H84 Cal Factor: Characterized by
e
standards lab to: ± 0.6%
Input Connector: Type-N (m)
Recommended
Al ternative
Keysight Model
Keysight Model
a
Number
or Vendor
Number
Model Number
33250A
33120A
(required for Channel
(Option 001)
Match (Option BBA)
(cannot be used
test)
for CXA
frequency
response
adjustment)
53132A
53131A
3458A
N1914A
E4419A/B
N1912A
8482A
N8482A
(Option H84)
N8482A CFT
8482A
(Non H84
sensors will
increase
measurement
uncertainty)
N9030B PXA Signal Analyzer Service Guide
b
Use
a
A, P
A, P, T
A, P, T
A, P
A, P
f
f
f