Calibrator Amplitude Accuracy Test - HP 8566B Adjustment Manual

Spectrum analyzer performance tests and adjustments manua
Hide thumbs Also See for HP 8566B:
Table of Contents

Advertisement

CAL OUTPUT Adjustment
-10 dBm f0.3 dB; 100 MHz
The output level of the calibrator signal is measured with a power
meter.
SPECTRUM ANALYZER
Figure 2-13. Calibrator Amplitude Accuracy
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..HP436 A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..HP8481 A
Adapter, Type N (f) to BNC (m) . . . . . . . . . . . . . . . . . . . . . . . . . . . .1250-1477
1. Connect equipment as shown in Figure 2-13.
2. Measure output level of the CAL OUTPUT signal. The value should
be - 10.0 dBm f0.3 dB.

9. Calibrator Amplitude Accuracy Test

POWER METER
POWER
SENSOR
Performance Tests 2-31
Setup