11-4
Status Structure

Clearing registers and queues

When the Model 2750 is turned on, the bits of all registers in the status structure are
cleared (reset to 0), and the two queues are empty. Commands to reset the event and event
enable registers, and the Error Queue are listed in
mands, any enable register can be reset by sending the 0 parameter value with the individ-
ual command to program the register.
NOTE
Table 11-1
Common and SCPI commands to reset registers and clear queues
*CLS
STATus:PRESet
*CLS
STATus:QUEue:CLEar
SYSTem:CLEar
Notes:
1. The Standard Event Enable Register is not reset by STATus:PRESet or *CLS. Send the 0 parameter value
with *ESE to reset all bits of that enable register to 0 (see
page
2. STATus:PRESet has no effect on the Error Queue.
3. Use either of the two clear commands to clear the Error Queue.
SYSTem:PRESet and *RST have no effect on status structure registers and
queues.
Commands
Reset registers
Clear error queue
11-9).
Model 2750 Multimeter/Switch System User's Manual
Table
Description
Reset all bits of the following event registers to 0:
Standard Event Register
Operation Event Register
Measurement Event Register
Questionable Event Register
Reset all bits of the following enable registers to 0:
Operation Event Enable Register
Measurement Event Enable Register
Questionable Event Enable Register
Clear all messages from Error Queue
Clear messages from Error Queue
Clear messages from Error Queue
"Status byte and service request commands,"
11-1. In addition to these com-
Notes
1
1
2
3
3