Model 2750 Multimeter/Switch System User's Manual
If the instrument displays the "OVERFLOW" message on a particular range, select a
higher range until an on-range reading is displayed. Use the lowest range possible without
causing an overflow to assure best accuracy and resolution.

Auto ranging

To enable auto range, press the AUTO key. The AUTO annunciator turns on when auto
ranging is selected. While auto ranging is enabled, the instrument automatically selects the
best range to measure the applied signal. Auto ranging should not be used when optimum
speed is required. Note that the AUTO key has no effect on temperature (TEMP).
Up-ranging occurs at 120% of range. The Model 2750 will down-range when the reading
is <10% of nominal range.
To disable auto ranging, press AUTO. This will leave the instrument on the present range.
You can also disable auto ranging by pressing the
may occur.

Scanning

When a simple scan is configured, the present function and range setting will apply to all
channels in the scan. When an advanced scan is configured, each channel can have its own
unique range setting. Details to configure and run a scan are provided in
For remote programming, the <clist> parameter is used to configure channels for a scan.
Range, Digits, Rate, Bandwidth, and Filter
Δ
or
key, however a range change
4-3
Section
7.