9-14
Limits and Digital I/O
Table 9-2 (continued)
Limits and digital I/O commands
Commands
Digital output commands
CALCulate3:OUTPut:LSENse <name>
CALCulate3:OUTPut:[STATe] <b>
CALCulate3:OUTPut:PULSe:TIMe
<NRf>
CALCulate3:OUTPut:PULSe[:STATe] <b> Enable/disable pulse output; <b> = ON or OFF. OFF
CALCulate3:MLIMit:LATChed <b>
Channel list parameter:
<clist> = (@SCH)
where: S
= Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
*The <clist> parameter is used to configure one or more channels for a scan.
NOTE
Reference
a.
NOTE
*
Set logic sense; <name> = AHIGh or ALOW.
Enable/disable digital outputs; <b> = ON or
OFF.
Set output pulse time in secs; <NRf> = 0.001
to 99999.999.
Enable/disable master limit latch.
When measurements are performed, the readings are fed to other enabled oper-
ations, including Limits.
and the commands used to read the result of limit tests.
CALCulate3:LIMit1:STATe <b> [, <clist>]
CALCulate3:LIMit2:STATe <b> [, <clist>]
Unlike front panel operation, Limit 1 and Limit 2 can be controlled (on/off) sepa-
rately for remote programming. The front panel limit indicators are affected as
follows:
Limit 1 enabled — The front panel HIGH/IN/LOW indicators work the same as
they do for front panel operation.
Limit 1 disabled and Limit 2 enabled — The status indicators pertain to Limit 2.
When the reading is within Limit 2, the message "I2" is displayed. When the read-
ing reaches or exceeds the high or low limit, the HIGH or LOW annunciator will
turn on, and the number "2" will be displayed.
When limits are disabled from the front panel, both Limit 1 and Limit 2 disable
for remote operation.