7-4
Scanning
closed channels, (2) close the paired channels, and then (3) perform the 4-wire measure-
ment. The last scanned channel pair opens.
NOTE
Calculations using channel pairs — Ratio and channel average performs measurements
on two channels and then calculates (and displays) the result. Therefore, these 2-channel
calculations also use paired channels. The scan process is to (1) open any closed channels,
(2) close the primary (displayed) channel and perform a measurement, (3) open the pri-
mary channel, (4) close the paired channel and perform a measurement, (5) calculate and
display the result, and finally, (6) open the paired channel.
NOTE

Trigger models

NOTE
Block diagrams, known as trigger models, are used to show the two fundamental scan
functions: STEP or SCAN. These two scan functions are enabled by the STEP and SCAN
keys, respectively. The trigger models for scanning are shown in
Figure
NOTE
For the Model 7700 switching module, primary channels 1 through 10 are
paired to channels 11 through 20. Channel 1 is paired to channel 11, channel 2
is paired to channel 12, channel 3 is paired to channel 13, and so on.
When scanning, the displayed channel number (i.e., 101) is not necessarily the
channel that is presently closed:
If both a reading AND a scan channel are displayed, the reading and annuncia-
tors pertain to that channel, but that channel is no longer closed. The next chan-
nel in the scan list is the one that is now closed. Therefore, the reading (and
annunciators) pertain to the channel and does not necessarily indicate the
present state of the Model 2750.
If the display is blanked (-------), the displayed channel is closed and has not
been measured.
The following information on trigger model operations apply specifically to
front panel operation.
7-2.
The trigger model in
programming — scanning," page
remote scanning.
For the following discussion, refer to
Figure 7-2
for SCAN operation.
Model 2750 Multimeter/Switch System User's Manual
Figure 7-2
also applies for bus operation. See
7-25, for differences between front panel and
Figure 7-1
Figure 7-1
and
"Remote
for STEP operation, and