The Trigger Link connections for this test system are shown in
the Model 2750 is connected to Trigger Link (either IN or OUT) of the Model 7002. Note
that with the default trigger settings on the Model 7002, line #1 is an input and line #2 is
an output. This complements the trigger lines on the Model 2750.
For this example, the Models 2750 and 7002 are configured as follows:
Model 2750
Factory defaults restored (accessed from SHIFT-SETUP).
External triggers (accessed from EX TRIG).
Buffer enabled and set to store 400 readings.
Model 7002
Factory defaults restored
Scan list = 1!1-1!400
Number of scans = 1
Channel spacing = TrigLink
Figure 8-7
Trigger link connections
Model 7002
WARNING:
INTERCONNECTION, INSTALLATION AND REMOVAL OF CARDS BY QUALIFIED SERVICE PERSONNEL ONLY .
CARD
CARD
CARD
CARD
1
2
3
4
WARNING:
NO INTERNAL OPERATOR SERVICEABLE PARTS, SERVICE BY QUALIFIED PERSONNEL
ONLY .
Trigger Link
CARD
CARD
CARD
CARD
CARD
5
6
7
8
9
!
DIGITAL I/O
TRIG. LINK
RS-232
IEEE-488
Trigger Link
I
CARD
N
10
TRIGGER
LINK
DIGITAL
O
I/O
U
T
A1
A2
A3
A4
A5
CHANNEL
READY
Trigger Link
EXTERNAL
IEEE-
TRIGGER
488
Cable (8501)
LINE
RATING
90-260
V
47-
440
85
VA
Hz
MAX
MADE IN USA
Model 2750
SLT
1
MADE IN
U.S.A.
SLT
2
CAT I
Triggering
Figure
8-7. Trigger Link of
KEITHLEY
SLT
3
SLOT COVER
KEITHLEY
SLT
4
SLOT COVER
!
KEITHLEY
SLT
5
SLOT COVER
8-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents