Keithley 2750 User Manual page 44

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System User's Manual
The following examples demonstrate how to include the function name in the command
string for configuration commands.
VOLT:RANG 10
RES:RANG:AUTO ON
CURR:DIG 4.5
NOTE
Exercise 1 — Basic DMM measurements
The exercise in
buffer.
Table 1-6
Exercise 1 — Measure AC volts - store readings in buffer
Front panel operation
1
For front panel operation, proceed to step 2.
For remote programming, clear the buffer
2
2
Restore defaults
Press SHIFT > press SETUP > select RESTORE: FACT.
3
Select ACV function:
Press ACV.
4
Select 10V range:
Press RANGE
6
Store 15 readings in buffer
Press STORE > set for 000015 RDGS > press ENTER.
7
Recall buffer readings
Press RECALL > use edit keys to display readings. Press EXIT to exit
recall mode.
1. To avoid problems with remote programming, it is good practice to routinely clear the buffer (
of a program that performs multiple measurements (
the buffer.
2. FACTory defaults place the instrument in a continuous measurement mode. *RST places the instrument in a non-
continuous measurement mode.
3. READ? triggers and returns 15 readings. These 15 readings are automatically stored in the buffer. See Exercise 4 and 5 for
more information on the READ? command.
4. Statistics for buffer readings are also stored in the buffer. For remote programming,
ings that were stored. It does not return buffer statistics.
(see
Section 6
for details).
See
Section 4
for details on setting range, digits, rate, bandwidth, and filter.
Table 1-6
measures ACV on the 10V range and stores 15 readings in the
:
Δ
to display "RANGE: 10V".
3
:
4
:
'
Select 10V range for DCV.
'
Enable auto range for W2.
'
Set DCI for 4½- digit resolution.
1
:
). Restoring *RST or FACTory defaults does not clear
SAMP:COUN >1
commands are used to calculate and return buffer statistics
CALC2
Getting Started
Command sequence
TRAC:CLE
*RST
FUNC 'VOLT:AC'
VOLT:AC:RANG 10
SAMP:COUN 15
READ?
CALC1:DATA?
) at the beginning
TRAC:CLE
only returns the read-
CALC1:DATA?
1-27

Hide quick links:

Advertisement

Table of Contents
loading