Model 2750 Multimeter/Switch System User's Manual
Table 7-1 (continued)
Scanning commands
Commands
Trigger commands
TRIGger:SOURce <name>
TRIGger:TIMer <n>
TRIGger:COUNt <NRf>
TRIGger:DELay <n>
TRIGger:DELay:AUTO <b>
SAMPle:COUNt <NRf>
SAMPle:COUNt?
INITiate:CONTinuous <b>
INITiate
READ?
Buffer commands
TRACe:DATA?
TRACe:CLEar
Channel list parameter:
<clist> = (@SCH)
where: S
= Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Notes:
1. Not affected by *RST and SYSTem:PRESet. Front panel factory default is OFF.
2. *RST sets count to 1, and SYSTem:PRESet sets count to INFinity.
3. The default value depends on which switching module is installed.
Description
Select control source; <name> = IMMediate, TIMer,
MANual, BUS, or EXTernal.
Set timer interval in sec; <n> = 0.001 to 999999.999.
Set trigger count; <NRf> = 1 to 110000, or INFinity.
Set delay in sec; <n> = 0 to 999999.999.
Enable/disable auto delay; <b> = ON or OFF.
Set sample count; <NRf> = 1 to 110000.
Query sample count.
Enable/disable continuous initiation; <b> = ON or OFF.
Initiate one scan cycle.
Initiate one scan cycle and request sample readings.
Read buffer readings.
Clear buffer.
Scanning
7-29
Default Ref
IMM
0.1
(Note 2) e
0
ON
1
e
f
f
f

Hide quick links: