Remote Programming - Ratio And Channel Average - Keithley 2750 User Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System User's Manual
Remote programming — ratio and channel average
Ratio and channel average commands
The ratio and channel average are listed in
the table.
NOTE
Table 5-3
Ratio and channel average commands
Commands*
[SENSe[1]]
:RATio[:STATe] <b> [, <clist>]
:RATio:DELay <NRf> [, <clist>]
:CAVerage[:STATe] <b> [, <clist>]
:CAVerage:DELay <NRf> [, <clist>]
Channel list parameter:
<clist> = (@SCH)
where: S
= Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
*The <clist> parameter is used to configure one or more channels for a scan.
Queries are not included in
Table
15-1.
Optional root command.
Enable/disable ratio; <b> = ON or OFF.
Set delay (in secs); <NRf> = 0 to 99999.999.
Enable/disable channel average; <b> = ON or OFF.
Set delay (in secs); <NRf> = 0 to 99999.999.
Rel, Math, Ratio, Channel Average, dB
Table
5-3. Details on these commands follow
Table
5-3. All the math commands are provided in
Description
5-19
Def
OFF
0.5
OFF
0.5