4-6
Range, Digits, Rate, Bandwidth, and Filter
Remote programming — digits
Digits commands
The commands to control display resolution (digits) are listed in
information on these commands follow the table.
NOTE
Table 4-3
Digits commands
Commands*
[SENSe[1]]
:VOLTage[:DC]:DIGits <n> [, <clist>]
:VOLTage:AC:DIGits <n> [, <clist>]
:CURRent[:DC]:DIGits <n> [, <clist>]
:CURRent:AC:DIGits <n> [, <clist>]
:RESistance:DIGits <n> [, <clist>]
:FRESistance:DIGits <n> [, <clist>]
:TEMPerature:DIGits <n> [, <clist>]
:FREQuency:DIGits <n> [, <clist>]
:PERiod:DIGits <n> [, <clist>]
Channel list parameter:
<clist> = (@SCH)
where: S
= Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
* The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the function
specified by the digits command. If not, a conflict error (-221) will occur. For example, VOLTage:AC:DIGits 4.5, (@101) is only
valid if scan channel 101 is set for the ACV function.
Query commands are not included in
subsystem are provided in
Optional root command.
Model 2750 Multimeter/Switch System User's Manual
Table
4-3. All commands for the SENSe
Table
15-5.
Description
Set # of digits for DCV; <n> = 4 to 7.
Set # of digits for ACV; <n> = 4 to 7.
Set # of digits for DC1; <n> = 4 to 7.
Set # of digits for ACI; <n> = 4 to 7.
Set # of digits for Ω2; <n> = 4 to 7.
Set # of digits for Ω4; <n> = 4 to 7.
Set # of digits for TEMP; <n> = 4 to 7.
Set # of digits for FREQ; <n> = 4 to 7.
Set # of digits for PERIOD; <n> = 4 to 7.
Table
4-3. Additional
Default
7
6
7
6
7
7
6
7
7