Monitor Channel - Keithley 2750 User Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

7-18
Scanning
NOTE
4.
For remote programming, the TRIGer:DELay <NRf> and TRIGger:DELay:AUTO <b>
commands are used to set the delay. See
NOTE

Monitor channel

While in the normal measurement state, a scan list channel can be used to monitor read-
ings. When a channel is selected to be the monitor, it will assume the setup of the scan list
channel.
NOTE
When a scan is started, the first channel in the scan list will be briefly displayed. While the
scan is in progress, the display will only show the reading(s) for the monitor channel.
After the last channel is scanned, the scan will disable with the monitor channel closed.
Monitor can be used with limit testing to trigger the start of a scan. When the monitor
detects that a set reading limit has been reached, the scan is triggered to start. The detailed
procedure to perform a monitor scan is provided in
scan," page
NOTE
To set monitor, the instrument can be in the normal measurement state or enabled while in
the advanced scan menu. There are two methods to select a monitor channel: method 1
selects it while a channel is closed, and method 2 selects it with no channels closed.
NOTE
Pressing the AUTO key sets the manual delay to 0.001 sec.
With the desired manual delay displayed, press ENTER.
The delay for ratio and channel average can only be set using remote program-
ming
(Table
5-3).
If you change the setup while a monitor channel is closed, that setup will be cop-
ied to that channel in the scan list.
7-35.
An overflow reading ("OVRFLW" message displayed) is interpreted by the
Model 2750 as a positive reading, even if the input signal is negative. This could
inadvertently trigger a monitor scan (see
scan," page
7-35.
The monitor channel must be a channel that is in the scan list. If the monitor
channel is removed from the scan list, the lowest channel in the scan list will
become the monitor channel.
The monitor channel must be a channel that is in the scan list.
Model 2750 Multimeter/Switch System User's Manual
Table 7-1
for details.
"(see Scan operation)
"(see Scan operation)
Monitor
Monitor

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents