Setting Delay - Keithley 2750 User Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System User's Manual
Step 3: Enable immediate scan
The present state of immediate scan (IMM SCAN) is displayed; Y (yes, which is the fac-
tory and *RST default) or N (no). With immediate scan enabled, the scan will start when
you press the STEP or SCAN key. Use the
press ENTER.
NOTE
Step 4: Timer controlled scan
The present state of the TIMER will be displayed; OFF or ON. If you do not wish to use
the timer, use the
default) and press ENTER.
To use the timer, use the
timer interval is displayed in the hour/minute/second format. The timer can be set from
0.001 sec (00H:00M:00.001S) to 99 hr, 99 min, 99.999 sec (99H:99M:99.999S). Note that
pressing the AUTO key sets the timer to 0.001 sec. With the desired interval displayed,
press ENTER.
Step 5: Set the reading count
The displayed reading count (RDG CT) sets the number channels to scan (STEP) or the
number of scans to run (SCAN). You can change the reading count to any finite value from
2 to 110000, or you can select infinite (continuous scanning). To select infinite, set the
reading count to 00000 to display INF. With the desired reading count setting displayed,
press ENTER. The instrument returns to the normal measurement mode.
See

Setting delay

As shown in
With auto delay selected, the delay period depends on function and range
With manual delay selected, the delay period can be set from 0 secs to 99 hrs, 99 mins,
99.999 secs.
Perform the following steps to set auto or manual delay:
1.
2.
3.
Disable immediate scan (IMM SCAN: N) when you wish to use a monitored
reading limit to trigger the start of the scan. This technique to start a scan is
covered later in this section. See
Δ
or
key to display TIMER? OFF (which is the factory and *RST
Δ
"Trigger models," page
Figure 7-1
and
With the instrument in the normal measurement display state, press SHIFT and
then DELAY.
Δ
Press
or
to display AUTO (auto delay) or MAN (manual delay) and press
ENTER.
If you selected MAN, you will be prompted to set the delay in the hour/minute/sec-
ond time format. Use the
Δ
or
"Scan operation," page
or
key to display TIMER? ON and press ENTER. The
7-4, for details on reading count.
Figure
7-2, a Delay (Auto or Manual) can be set by the user.
Δ
,
,
, and
Scanning
key to display IMM SCAN: Y, and
7-21.
(Table
keys to set the delay.
7-17
8-1).