Setting Digital Output - Keithley 2750 User Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System User's Manual

Setting digital output

The OUTPUT menu (shown in
Menu items for the digital output include:
NOTE
Table 9-1
OUTPUT menu
Menu item
DOUTPUT
PULSE
LSENSE
BEEP
MASTR LATCH Y or N
DOUTPUT — Use to enable (ON) or disable (OFF) the digital outputs.
PULSE — Use to enable (YES) or disable (NO) the pulse option for the digital
outputs.
The factory default pulse time is 2ms maximum. Using remote programming,
pulse time can be set from 0.001 to 99999.999 sec. It cannot be set from the front
panel.
LSENSE — Use to select the logic sense; active HIGH or active LOW. With active
high selected, an output will be at approximately +5V when a reading is at or
exceeds the limit. Conversely, with active low selected, an output will be at 0V
when a reading reaches or exceeds the limit.
MASTR LATCH — Use to enable (Y) or disable (N) the master limit latch. When
enabled, the master limit remains latched when a reading limit is reached or
exceeded. When disabled, the master limit line releases immediately when the
reading is inside all four limits.
ON or OFF
YES or NO
HIGH or LOW
NEVER, INSIDE, or OUTSIDE Set beeper for limits
Table
9-1) is used to control and configure digital outputs.
Setting
Limits and Digital I/O
Description
Enable/disable digital outputs.
Enable/disable digital pulse
output.
Select logic sense.
(see
"Limits," page
9-2, for
details).
Enable/disable master limit latch.
9-11