9-10
Limits and Digital I/O
Source mode — logic control
The digital outputs can be used as logic inputs to active TTL, low-power TTL, or CMOS
inputs. For this mode of operation, the output lines can source up to 200µA.
CAUTION
Figure 9-5
put line is pulled high, the transistor will turn off (transistor switch open) to provide a reli-
able logic high output (>3.75V). When the output line goes low, the transistor turns on
(transistor switch closed) to route current to digital ground. As a result, a low logic output
(0V) is provided at the output.
If the second input (B) of the NAND gate is connected to another output line of the port,
the output of the NAND gate will go to logic 0 when both digital outputs are set high.
Figure 9-5
NAND gate control
Control
Line
Each output line can source up to 200µA. Exceeding 200µA may cause
damage to Model 2750 that is not covered by the warranty.
shows how to connect a logic device to one of the output lines. When the out-
Model 2750
+5V
4.75kΩ
Pull Up Resistor
Model 2750 Multimeter/Switch System User's Manual
B
Digital
A
Output
Pin 9
Logic Device
NAND