10-6
Remote Operations
Figure 10-2
Figure 10-2
IEEE-488 connections
Instrument
To avoid possible mechanical damage, stack no more than three connectors on any one
unit.
NOTE
To connect the Model 2750 to the IEEE-488 bus, follow these steps:
1.
2.
3.
4.
NOTE
shows a typical connecting scheme for a multi-unit test system.
Instrument
To minimize interference caused by electromagnetic radiation, use only shielded
IEEE-488 cables. Available shielded cables from Keithley are Models 7007-1
and 7007-2.
Line up the cable connector with the connector located on the rear panel. The con-
nector is designed so it will fit only one way.
IEEE-488 connector.
Tighten the screws securely, making sure not to overtighten them.
Connect any additional connectors from other instruments as required for your
application.
Make sure the other end of the cable is properly connected to the controller. Most
controllers are equipped with an IEEE-488 style connector, but a few may require a
different type of connecting cable. See your controller's instruction manual for
information about properly connecting to the IEEE-488 bus.
You can only have 15 devices connected to an IEEE-488 bus, including the con-
troller. The maximum cable length is either 20 meters or two meters multiplied
by the number of devices, whichever is less. Not observing these limits may
cause erratic bus operation.