Model 2750 Multimeter/Switch System User's Manual
Reference
a.
FUNCtion <name> [, <clist>]
Note that the <name> parameters in the table are enclosed in single quotes
(' '). However, double quotes (" ") can instead be used. For example:
FUNC 'VOLT:AC' = FUNC "VOLT:AC"
Scan configuration — When using the <clist> command to configure a scan
channel, the scan channel must first be set to the appropriate function before
sending other commands to configure it. For example, to set scan channel 101
to use offset compensated ohms, the following command sequence would be
sent:
FUNC 'FRES',
(@101)
FRES:OCOM
ON, (@101)
If scan channel 101 was not first set for the Ω4 function, error +700 (Invalid
function in scanlist) will occur when trying to enable offset compensated ohms
for that channel. Details on scanning are provided in
b.
FRESistance:OCOMpensated <b> [, <clist>]
FRESistance:DCIRcuit <b> [, <clist>]
The instrument does not have to be on the Ω4 function to enable offset-
compensated ohms and/or dry circuit ohms. When Ω4 is selected, offset-
compensated ohms and/or dry circuit ohms will be enabled.
When using the <clist> parameter to configure a scan channel for offset-
compensated ohms, that channel must already be set for Ω4. If set for another
function, a settings conflict (-221) will occur.
c.
TEMPerature:TCouple:RJUNction:RSELect <name> [, <clist>]
To use the EXTERnal reference junction, the scan channel (101, 201, 301,
401, or 501) to be used for the measurement must already be configured to use
a thermistor or 4-wire RTD transducer. Otherwise, a settings conflict error
(-221) will occur. The TEMPerature:TRANsducer command is used to select
the transducer.
NOTE
The following command can instead be used to select the reference
junction:
TEMPerature:RJUNction:RSELect <name> [, <clist>]
d.
TEMPerature:TCouple:RJUNction:SIMulated <n> [, <clist>]
The units for the simulated reference temperature depend on the present tem-
perature measurement units as set by UNIT:TEMPerature (see Ref h).
NOTE
The following command can instead be used to set the simulated ref-
erence temperature:
TEMPerature:RJUNction:SIMulated <n> [, <clist>]
Basic DMM Operation
' Set scan channel 101 to Ω4 function.
' Enable offset compensated ohms for scan
' channel 101.
3-45
Section
7.