1-26
Getting Started
NOTE
Basic DMM measurements — front panel inputs
NOTE
The Model 2750 is shipped from the factory to power-up to factory defaults. The instru-
ment powers up to a setup that continuously measures DC volts. Some of the default set-
tings for the DCV function include auto range enabled, 6½-digit resolution, filter enabled,
and slow reading rate. These settings provide a good starting point and in many cases, do
not need to be changed.
"Starting-point" default settings are also provided for the other measurement functions.
Therefore, to perform basic measurements, simply select the desired function, and
"tweak" the setup (range, rate, filter, digits, etc.) as required.
For remote programming, the instrument is typically used in a non-continuous measure-
ment mode. In this mode, the user (via remote command programming) specifies the num-
ber of measurements to perform. *RST defaults place the instrument in a non-continuous
measurement mode. Most of the other settings for factory and *RST defaults are the same.
For remote programming, the following command is used to select function.
NOTE
[SENSe[1]]:FUNCtion <func>
Each function can have its own unique setup configuration (i.e., range, digits, speed, etc.).
For example, the following command words select range and digits:
RANGe[:UPPer] <n>
RANGe:AUTO <b>
DIGits
When using the front panel input terminals, the INPUT switch must be in the
"F" (out) position. The INPUT switch is located on the right side of the front
panel near the input terminals. When using a switching module, the switch must
be in the "R" (in) position.
See
Section 3
for details on basic DMM operation.
Items in brackets ([]) are optional and do not need to be included. Upper case
characters are required. Lower case characters are optional and need not be
included.
<func>
=
'VOLTage[:DC]'
'VOLTage:AC'
'CURRent[:DC]'
'CURRent:AC'
'RESistance'
'FRESistance'
'FREQuency'
'PERiod'
'TEMPerature'
' Specify expected reading.
' Enable (ON) or disable (OFF) auto range.
' Set display resolution; 3.5, 4.5, 5.5 or 6.5 (digits).
Model 2750 Multimeter/Switch System User's Manual
'
Select measurement function.
DCV
ACV
DCI
ACI
Ω2
Ω4
FREQ
PERIOD
TEMP