Keithley 2750 User Manual page 337

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

13-6
SCPI Signal Oriented Commands
NOTE
NOTE
If the instrument is in the idle state, INITiate takes the instrument out of
the idle state. If continuous initiation is enabled, (INITiate:CONTinuous
ON), then the INITiate command generates error -213 (init ignored).
If continuous initiation is enabled, the READ? command will generate
error -213 (init ignored). It can be disabled by sending INITiate:CONTinuous
OFF. The *RST command can also be used to disable continuous initiation. It
also places the Model 2750 in the "one-shot" measurement mode:
*RST
' Disable continuous initiation and place 2750 in "one-shot" mode.
READ?
' Trigger and return one reading.
When readings are stored in the buffer by the TRACe command (or by front
panel data store operation), INIT and multi-sample READ? queries are locked
out. With readings in the buffer that were stored in that manner, you cannot use
the INIT or READ? command if sample count is >1 (error -225, out of memory).
Buffer operation is covered in
The buffer of the Model 2750 is nonvolatile. Therefore, readings stored in the
buffer are not lost when the instrument is turned off, or when *RST or
SYSTem:PRESet is sent. When writing test programs that perform multi-sample
measurements (SAMPle:COUNTt >1), you may want to add the TRACe:CLEar
command at the beginning to clear the buffer. However, be careful not to inad-
vertently clear stored readings that are needed.
Model 2750 Multimeter/Switch System User's Manual