Keithley 2750 User Manual page 135

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

3-44
Basic DMM Operation
Table 3-8 (continued)
Basic measurement commands
Commands
PERIOD function
:PERiod:THReshold:VOLTage:RANGe
<n> [, <clist>]
:PERiod:APERture <NRf> [, <clist>]
CONT function
[SENSe[1]]
:CONTinuity:THReshold <NRf>
SYSTem:BEEPer <b>
Set temperature measurement units
UNIT:TEMPerature <name>
Trigger and retrieve readings
INITiate:CONTinuous <b>
INITiate
[SENSe[1]]
:DATA[:LATest]?
:DATA:FRESh?
FETCh?
READ?
Channel list parameter:
<clist> = (@SCH)
where:
S = Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
1. The <clist> parameter is used to configure one or more channels for a scan.
2. When using multiple channel operation (ROUT:MULT command) to connect a switching module channel to the DMM for
thermocouple temperature measurements, the SIMulated reference junction will be used if the INTernal or EXTernal reference
junction is selected. The "ERR" annunciator will turn on to indicate that the integrity of the temperature reading is questionable.
3. With a Model 7700, 7706, or 7708 installed, the default sensor junction is Internal. Otherwise, the Simulated (23°C) junction is
selected.
4. The *RST default is OFF, and the SYSTem:PRESet default is ON.
1
Select threshold voltage range; <n> =
Set gate time for PERIOD measurements
Optional root command.
Set continuity threshold in ohms;
Enable/disable beeper; <b> = ON or OFF.
Set temperature units; <name> = C, CEL,
F, FAR, or K.
Enable/disable continuous initiation;
<b> = ON or OFF.
Trigger one or more measurements.
Optional root command.
Returns the last reading string.
Returns the last "fresh" reading string.
Return reading(s).
Trigger and return reading(s).
Model 2750 Multimeter/Switch System User's Manual
Description
0 to 1010.
in secs; <NRf> = 0.01 to 1.0.
<NRf> = 1 to 1000.
Default Ref
10
f
1.0
g
10
ON
h
i
Note 4

Hide quick links: