Configure:<Function> [<Rang>], [<Res>], [<Clist>] - Keithley 2750 User Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System User's Manual
CONFigure:<function> [<rang>], [<res>], [<clist>]
CONFigure:VOLTage[:DC] [<rang>], [<res>], [<clist>]
CONFigure:VOLTage:AC [<rang>], [<res>], [<clist>]
CONFigure:CURRent[:DC] [<rang>], [<res>], [<clist>]
CONFigure:CURRent:AC [<rang>], [<res>], [<clist>]
CONFigure:RESistance [<rang>], [<res>], [<clist>]
CONFigure:FRESistance [<rang>], [<res>], [<clist>]
CONFigure:FREQuency [<rang>], [<res>], [<clist>]
CONFigure:PERiod [<rang>], [<res>], [<clist>]
CONFigure:TEMPerature [<clist>]
CONFigure:CONTinuity [<clist>]
Parameters
<rang> = Range parameter for the specified function. For example, for
DCV, range parameter value 10 selects the 10V range. See
the "NOTES" that follow
<res> =
0.1
0.01
0.001
0.0001
0.00001
0.000001
The resolution of the <res> parameter value and the selected range
sets the number of display digits. As shown above, with the 100V
range selected and <res> = 0.1, a 100V reading will be displayed
as 100.0 V (3H digits).
The display will default to 3H digits when using parameter values
that attempt to set the display below 3H digits. For example, a 10V
reading using <res> = 0.1 for the 10V range is displayed as 10.00
V, not 10.0 V.
A command using parameter values that attempt to set the dis-
play above 7H digits is ignored, and generates error -221 (set-
tings conflict error).
The <res> parameter is ignored when a <clist> is included in
the command string. Resolution for the scanlist channel(s) is
determined by the present setting for the specified function and
by the present resolution setting for the specified function.
See the "NOTES" that follow
mation.
<clist> = Channel(s) in the scanlist to be configured. When the channel
list parameter (<clist>) is included, the present instrument set-
tings are not affected. Instead, the channel(s) in the <clist> for
the specified function is configured. See the "NOTES" that fol-
low
Table 13-1
SCPI Signal Oriented Commands
Configure DCV
Configure ACV
Configure DCI
Configure ACI
Configure Ω2
Configure Ω4
Configure FREQ
Configure PERIOD
Configure TEMP
Configure CONT
Table 13-1
i.e., 100.0 V (3H digits)
i.e., 10.00 V (3H digits)
i.e., 1.000 V (3H digits)
i.e., 1.0000 V (4H digits)
i.e., 1.00000 V (5H digits)
i.e., 1.000000 V (6H digits)
Table 13-1
for additional information.
13-3
for additional information.
for additional infor-