Remote Programming For Basic Measurements; Basic Measurement Commands - Keithley 2750 User Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

3-42
Basic DMM Operation

Remote programming for basic measurements

Basic measurement commands

NOTE
Commands to perform basic measurements are listed in
Table 3-8
Basic measurement commands
Commands
Select measurement function
[SENSe[1]]
:FUNCtion <name> [, <clist>]
DCV function
[SENSe[1]]
:VOLTage[:DC]:IDIVider <b>
Ω4 function
[SENSe[1]]
:FRESistance:OCOMpensated <b>
[, <clist>]
:FRESistance:DCIRcuit <b> [, <clist>]
When measurements are performed, the readings are fed to other enabled pro-
cessing operations.
Appendix D
the commands used to return the various processed readings.
1
Optional root command.
Select measurement function:
Optional root command.
Enable/disable DCV input divider;
Optional root command.
Enable/disable offset compensated ohms. OFF
Enable/disable dry circuit ohms.
Model 2750 Multimeter/Switch System User's Manual
explains
"Data flow (remote
Table
Description
<name> = 'VOLTage[:DC]',
'VOLTage:AC', 'CURRent[:DC]',
'CURRent:AC', 'RESistance',
'FRESistance', 'CONTinuity',
'FREQuency', or 'PERiod'.
Note: [:DC] is optional.
<b> = ON or OFF.
operation)" and
3-8.
Default Ref
VOLT
OFF
OFF
a
b
b