Keithley 2750 User Manual page 360

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System User's Manual
Table 15-5 (continued)
SENSe command summary
Command
[SENSe[1]]
:RESistance
:APERture <NRf>
[, <clist>]
:APERture? [<clist>]
:NPLCycles <n>[, <clist>]
:NPLCycles? [<clist>]
:RANGe
[:UPPer] <n> [, <clist>]
[:UPPer]? [<clist>]
:AUTO <b> [, <clist>]
:AUTO? [<clist>]
:DIGits <n> [, <clist>]
:DIGits? [<clist>]
:REFerence <n> [, <clist>]
:STATe <b> [, <clist>]
:STATe? [<clist>]
:ACQuire [, <clist>]
:REFerence? [<clist>]
:AVERage
:TCONtrol <name>
:TCONtrol?
:WINDow <NRf>
:WINDow?
:COUNt <n> [, <clist>]
:COUNt? [<clist>]
[:STATe] <b> [, <clist>]
[:STATe]? [<clist>]
Description
Path to configure resistance.
Set integration rate in seconds (60Hz,
1.67e-4 to 1; 50Hz, 2e-4 to 1).
Query aperture integration rate.
Set integration rate in line cycles (60Hz,
0.01 to 60; 50Hz, 0.01 to 50).
Query line cycle integration rate.
Path to set measurement range:
Select range (0 to 120e6).
Query range.
Enable or disable auto range.
Query state of auto range.
Specify measurement resolution (4 to 7).
Query resolution.
Specify reference (0 to 120e6).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Path to configure and control filter:
Select filter type: (MOVing or REPeat).
Query filter type.
Set filter window in % of range
(0 to 10).
Query filter window.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
SCPI Reference Tables
Default
parameter
Ref
Sec 3
(Note 2)
Sec 4
5.0
Sec 4
Sec 4
120e6
ON
7
Sec 4
0
Sec 5
OFF
Sec 4
(Note 3)
0.1
10
(Note 4)
15-13
SCPI

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents